2.2.23. ATOMIC ABSORPTION SPECTROMETRY
GENERAL PRINCIPLE
Atomic absorption is a process that occurs when a ground state-atom absorbs
electromagnetic radiation of a specific wavelength and is elevated to an
excited state. The atoms in the ground state absorb energy at their resonant
frequency and the electromagnetic radiation is attenuated due to resonance
absorption. The energy absorption is virtually a direct function of the number
of atoms present.
This chapter provides general information and defines the procedures used in
element determinations by atomic absorption spectrometry, either atomisation
by flame, by electrothermal vaporisation in a graphite furnace, by hydride
generation or by cold vapour technique for mercury.
Atomic absorption spectrometry is a technique for determining the
concentration of an element in a sample by measuring the absorption of
electromagnetic radiation by the atomic vapour of the element generated from
the sample. The determination is carried out at the wavelength of one of the
absorption (resonance) lines of the element concerned. The amount of
radiation absorbed is, according to the Lambert-Beer law, proportional to the
element concentration.
APPARATUS
This consists essentially of:
a source of radiation;
a sample introduction device;
a sample atomiser;
a monochromator or polychromator;
a detector;
a data-acquisition unit.
The apparatus is usually equipped with a background correction system.
Hollow-cathode lamps and electrodeless discharge lamps (EDL) are used as
radiation source. The emission of such lamps consists of a spectrum showing
very narrow lines with half-width of about 0.002 nm of the element being
determined.
There are 3 types of sample atomisers:
Flame technique
A flame atomiser is composed of a nebulisation system with a pneumatic
aerosol production accessory, a gas-flow regulation and a burner. Fuel-
oxidant mixtures are commonly used to produce a range of temperatures from
about 2000 K to 3000 K. Fuel gases include propane, hydrogen and acetylene;
air and nitrous oxide are used as oxidants. The configuration of the burner is
adapted to the gases used and the gas flow is adjustable. Samples are
nebulised, acidified water being the solvent of choice for preparing test and
reference solutions. Organic solvents may also be used if precautions are
taken to ensure that the solvent does not interfere with the stability of the
flame.
Electrothermal atomisation technique
An electrothermal atomiser is generally composed of a graphite tube furnace
and an electric power source. Electrothermal atomisation in a graphite tube
furnace atomises the entire sample and retains the atomic vapour in the light
path for an extended period. This improves the detection limit. Samples, liquid
as well as solid, are introduced directly into the graphite tube furnace, which is
heated in a programmed series of steps to dry the sample and remove major
matrix components by pyrolysis and to then atomise all of the analyte. The
furnace is cleaned using a final temperature higher than the atomisation
temperature. The flow of an inert gas during the pyrolysis step in the graphite
tube furnace allows a better performance of the subsequent atomisation
process.
Cold vapour and hydride technique
The atomic vapour may also be generated outside the spectrometer. This is
notably the case for the cold-vapour method for mercury or for certain
hydride-forming elements such as arsenic, antimony, bismuth, selenium and
tin. For mercury, atoms are generated by chemical reduction with stannous
chloride or sodium borohydride and the atomic vapour is swept by a stream of
an inert gas into a cold quartz cell mounted in the optical path of the
instrument. Hydrides thus generated are swept by an inert gas into a heated
cell in which they are dissociated into atoms.
INTERFERENCES
Chemical, physical, ionisation and spectral interferences are encountered in
atomic absorption measurements. Chemical interference is compensated by
addition of matrix modifiers, of releasing agents or by using high temperature
produced by a nitrous oxide-acetylene flame; the use of specific ionisation
buffers (for example, lanthanum and caesium) compensates for ionisation
interference; by dilution of the sample, through the method of standard
additions or by matrix matching, physical interference due to high salt content
or viscosity is eliminated. Spectral interference results from the overlapping of
resonance lines and can be avoided by using a different resonance line. The
use of Zeeman background correction also compensates for spectral
interference and interferences from molecular absorption, especially when
using the electrothermal atomisation technique. The use of multi-element
hollow-cathode lamps may also cause spectral interference. Specific or non-
specific absorption is measured in a spectral range defined by the band-width
selected by the monochromator (0.2-2 nm).
BACKGROUND CORRECTION
Scatter and background in the flame or the electrothermal atomisation
technique increase the measured absorbance values. Background absorption
covers a large range of wavelengths, whereas atomic absorption takes place
in a very narrow wavelength range of about 0.005-0.02 nm. Background
absorption can in principle be corrected by using a blank solution of exactly
the same composition as the sample, but without the specific element to be
determined, although this method is frequently impracticable. With the
electrothermal atomisation technique the pyrolysis temperature is to be
optimised to eliminate the matrix decomposition products causing background
absorption. Background correction can also be made by using 2 different light
sources, the hollow-cathode lamp that measures the total absorption (element
+ background) and a deuterium lamp with a continuum emission from which
the background absorption is measured. Background is corrected by
subtracting the deuterium lamp signal from the hollow-cathode lamp signal.
This method is limited in the spectral range on account of the spectra emitted
by a deuterium lamp from 190-400 nm. Background can also be measured by
taking readings at a non-absorbing line near the resonance line and then
subtracting the results from the measurement at the resonance line. Another
method for the correction of background absorption is the Zeeman effect
(based on the Zeeman splitting of the absorption line in a magnetic field). This
is particularly useful when the background absorption shows fine structure. It
permits an efficient background correction in the range of 185-900 nm.
CHOICE OF THE OPERATING CONDITIONS
After selecting the suitable wavelength and slit width for the specific element,
the need for the following has to be ascertained:
correction for non-specific background absorption,
chemical modifiers or ionisation buffers to be added to the sample as well as
to blank and reference solutions,
dilution of the sample to minimise, for example, physical interferences,
details of the temperature programme, preheating, drying, pyrolysis,
atomisation, post-atomisation with ramp and hold times,
inert gas flow,
matrix modifiers for electrothermal atomisation (furnace),
chemical reducing reagents for measurements of mercury or other hydride-
forming elements along with cold vapour cell or heating cell temperature,
specification of furnace design (tank, Lvov platform, etc).
METHOD
Use of plastic labware is recommended wherever possible. The preparation of
the sample may require a dissolution, a digestion (mostly microwave-assisted),
an ignition step or a combination thereof in order to clear up the sample matrix
and/or to remove carbon-containing material. If operating in an open system,
the ignition temperature should not exceed 600 °C, due to the volatility of
some metals, unless otherwise stated in the monograph.
Operate an atomic absorption spectrometer in accordance with the
manufacturers instructions at the prescribed wavelength. Introduce a blank
solution into the atomic generator and adjust the instrument reading so that it
indicates maximum transmission. The blank value may be determined by
using solvent to zero the apparatus. Introduce the most concentrated
reference solution and adjust the sensitivity to obtain a maximum absorbance
reading. Rinse in order to avoid contamination and memory effects. After
completing the analysis, rinse with water R or acidified water.
If a solid sampling technique is applied, full details of the procedure are
provided in the monograph.
Ensure that the concentrations to be determined fall preferably within the
linear part of the calibration curve. If this is not possible, the calibration plots
may also be curved and are then to be applied with appropriate calibration
software.
Determinations are made by comparison with reference solutions with known
concentrations of the element to be determined either by the method of direct
calibration (Method I) or the method of standard additions (Method II).
METHOD I - DIRECT CALIBRATION
For routine measurements 3 reference solutions and a blank solution are
prepared and examined.
Prepare the solution of the substance to be examined (test solution) as
prescribed in the monograph. Prepare not fewer than 3 reference solutions of
the element to be determined, the concentrations of which span the expected
value in the test solution. For assay purposes, optimal calibration levels are
between 0.7 and 1.3 times the expected content of the element to be
determined or the limit prescribed in the monograph. For purity determination,
calibration levels are the limit of detection and 1.2 times the limit specified for
the element to be determined. Any reagents used in the preparation of the
test solution are added to the reference and blank solutions at the same
concentration.
Introduce each of the solutions into the instrument using the same number of
replicates for each of the solutions to obtain a steady reading.
Calculation
. Prepare a calibration curve from the mean of the readings
obtained with the reference solutions by plotting the means as a function of
concentration. Determine the concentration of the element in the test solution
from the curve obtained.
METHOD II - STANDARD ADDITIONS
Add to at least 3 similar volumetric flasks equal volumes of the solution of the
substance to be examined (test solution) prepared as prescribed. Add to all
but 1 of the flasks progressively larger volumes of a reference solution
containing a known concentration of the element to be determined to produce
a series of solutions containing steadily increasing concentrations of that
element known to give responses in the linear part of the curve, if possible.
Dilute the contents of each flask to volume with solvent.
Introduce each of the solutions into the instrument, using the same number of
replicates for each of the solutions, to obtain a steady reading.
Calculation
. Calculate the linear equation of the graph using a least-squares
fit and derive from it the concentration of the element to be determined in the
test solution.
VALIDATION OF THE METHOD
Satisfactory performance of methods prescribed in monographs is verified at
suitable time intervals.
LINEARITY
Prepare and analyse not fewer than 4 reference solutions over the calibration
range and a blank solution. Perform not fewer than 5 replicates.
The calibration curve is calculated by least-square regression from all
measured data. The regression curve, the means, the measured data and the
confidence interval of the calibration curve are plotted. The operating method
is valid when:
the correlation coefficient is at least 0.99,
the residuals of each calibration level are randomly distributed around the
calibration curve.
Calculate the mean and relative standard deviation for the lowest and highest
calibration level.
When the ratio of the estimated standard deviation of the lowest and the
highest calibration level is less than 0.5 or greater than 2.0, a more precise
estimation of the calibration curve may be obtained using weighted linear
regression. Both linear and quadratic weighting functions are applied to the
data to find the most appropriate weighting function to be employed. If the
means compared to the calibration curve show a deviation from linearity, two-
dimensional linear regression is used.
ACCURACY
Verify the accuracy preferably by using a certified reference material (CRM).
Where this is not possible, perform a test for recovery.
Recovery
. For assay determinations a recovery of 90 per cent to 110 per cent
is to be obtained. For other determinations, for example, for trace element
determination the test is not valid if recovery is outside of the range 80 per
cent to 120 per cent at the theoretical value. Recovery may be determined on
a suitable reference solution (matrix solution) which is spiked with a known
quantity of analyte (middle concentration of the calibration range).
REPEATABILITY
The repeatability is not greater than 3 per cent for an assay and not greater
than 5 per cent for an impurity test.
LIMIT OF QUANTIFICATION
Verify that the limit of quantification (for example, determined using the 10 σ approach) is below the value to be measured.